免费久久无码精品一区二区,国产AV一区二区最新精品无删减,欧美久久男人的天堂,免费国产小视频国产

產(chǎn)品中心

一步法人視黃醇結合蛋白(RBP)酶聯(lián)免疫吸附檢測試劑盒
ELK060ES
規(guī)格: 價格:
64T ¥1960.00
112T ¥2800.00

Overview

Product name: EasyStep Human RBP(Retinol Binding Protein) ELISA Kit
Reactivity: Human
Alternative Names: CRABP-I; CRBP; CRBP1; CRBPI; RBPC; hCRBP1
Assay Type: Sandwich
Sensitivity: 0.34 ng/mL
Standard: 100 ng/mL
Detection Range: 1.56-100 ng/mL
Sample Type: Serum, plasma
Assay Length: 1.5h
Research Area: Cancer; Diabetes Research; Lipid and Metabolism
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human RBP, and the Human RBP standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human RBP. After TMB substrate solution is added, only those wells that contain Human RBP and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human RBP in the samples is then determined by comparing the OD of the samples to the standard curve.

標準曲線

Concentration (ng/mL) OD Corrected OD
100.00 2.195 2.124
50.00 1.602 1.531
25.00 1.216 1.145
12.50 0.741 0.670
6.25 0.541 0.470
3.13 0.315 0.244
1.57 0.232 0.161
0.00 0.071 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant RBP and the recovery rates were calculated by comparing the measured value to the expected amount of RBP in samples.
Matrix Recovery range Average
serum(n=5) 85-97% 91%
EDTA plasma(n=5) 78-92% 85%
Heparin plasma(n=5) 92-105% 98%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of RBP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-102% 87-96% 78-92% 85-97%
EDTA plasma(n=5) 98-105% 86-95% 94-103% 95-102%
Heparin plasma(n=5) 95-103% 88-96% 81-95% 91-102%