免费久久无码精品一区二区,国产AV一区二区最新精品无删减,欧美久久男人的天堂,免费国产小视频国产

產(chǎn)品中心

犬轉(zhuǎn)鐵蛋白(TRF)酶聯(lián)免疫吸附檢測(cè)試劑盒
ELK1356
規(guī)格: 價(jià)格:
48T ¥1680.00
96T ¥2400.00

Overview

Product name: Dog TRF(Transferrin) ELISA Kit
Reactivity: Dog
Alternative Names: TF; Siderophilin; Serotransferrin; Beta-1 metal-binding globulin
Assay Type: Sandwich
Sensitivity: 1.12 ng/mL
Standard: 200 ng/mL
Detection Range: 3.13-200 ng/mL
Sample Type: Serum, plasma and other biological fluids.
Assay Length: 3.5h
Research Area: Tumor immunity;Infection immunity;Immune molecule;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Dog TRF. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Dog TRF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Dog TRF, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Dog TRF in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (ng/mL) OD Corrected OD
200.00 2.103 1.998
100.00 1.652 1.547
50.00 1.223 1.118
25.00 0.847 0.742
12.50 0.525 0.420
6.25 0.303 0.198
3.13 0.214 0.109
0.00 0.105 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant TRF and the recovery rates were calculated by comparing the measured value to the expected amount of TRF in samples.
Matrix Recovery range Average
serum(n=5) 85-97% 91%
EDTA plasma(n=5) 83-99% 95%
Heparin plasma(n=5) 85-97% 91%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TRF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-103% 79-92% 97-105% 89-97%
EDTA plasma(n=5) 85-94% 97-106% 82-95% 93-105%
Heparin plasma(n=5) 85-92% 83-96% 78-98% 90-99%