免费久久无码精品一区二区,国产AV一区二区最新精品无删减,欧美久久男人的天堂,免费国产小视频国产

產(chǎn)品中心

人表面活性物質(zhì)關(guān)聯(lián)蛋白B(SPB)酶聯(lián)免疫吸附檢測(cè)試劑盒
ELK1771
規(guī)格: 價(jià)格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human SPB(Surfactant Associated Protein B) ELISA Kit
Reactivity: Human
Alternative Names: SFTPB; PSPB; SFTB3; SFTP3; SP-B; Pulmonary Surfactant Protein B; 18 kDa pulmonary-surfactant protein; 6 kDa protein
Assay Type: Sandwich
Sensitivity: 0.6 ng/mL
Standard: 100 ng/mL
Detection Range: 1.57-100 ng/mL
Sample Type: serum, plasma, lung lavage fluid, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Tumor immunity;Infection immunity;Pulmonology;
Uniprot ID: P07988
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human SPB. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human SPB. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human SPB, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human SPB in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (ng/mL) OD Corrected OD
100.00 1.978 1.889
50.00 1.576 1.487
25.00 1.136 1.047
12.50 0.836 0.747
6.25 0.521 0.432
3.13 0.342 0.253
1.57 0.196 0.107
0.00 0.089 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant SPB and the recovery rates were calculated by comparing the measured value to the expected amount of SPB in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 92%
EDTA plasma(n=5) 84-96% 90%
Heparin plasma(n=5) 96-107% 101%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of SPB and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 90-98% 85-92% 83-96% 93-102%
EDTA plasma(n=5) 83-95% 88-102% 85-97% 80-92%
Heparin plasma(n=5) 98-107% 93-102% 80-93% 91-100%